Stainless Steel 2.0-mm Locking Compression Plate Osteosynthesis System for the Fixation of Comminuted Hand Fractures in Asian Adults
详细信息    查看全文
文摘

Objective

The aim of this retrospective study was to analyse the clinical outcome of the application of stainless steel 2.0-mm locking compression plate (LCP) system for the treatment of comminuted hand fractures in Asian adults.

Methods

Six patients who had comminuted hand fractures were treated by open reduction and internal fixation with the application of stainless steel 2.0-mm LCP (AO Compact Hand System; Synthes, Oberdorf, Switzerland) from December 2009 to October 2010. The total arc of motion of fingers, grip power, complications, and additional surgery were recorded.

Results

Three out of six patients eventually restored good hand functions in terms of the total arc of finger motion (>220¡ã) and grip power. The commonest complication was skin impingement in finger region by the implant (4 cases). Another common complication was restricted range of motion (3 cases). One patient had minimal degree of malrotation of his left little finger. Additional surgery was required in all the patients for implant removal (6 cases), tenolysis (3 cases), and capsulotomy (2 cases).

Conclusions

The stainless steel 2.0-mm LCP is useful for the fixation of unstable comminuted hand fractures, especially in metacarpal bones, because of its advantage of better stability, which allows more aggressive rehabilitation. However, its design is not very versatile and, therefore, limits its use in the finger region. Its bulkiness frequently causes implant impingement. The patients must be informed about the chance of implant removal later.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700