摘要
首先分析了绝压变送器的结构原理、使用环境、人为调试等方面影响产生零位漂移的现象;然后研究了不同量程的绝压变送器在进行零点调整时产生的线性误差;最后,针对小量程绝压变送器校准过程中由于设备真空度不足造成的线性差问题,提出了采用"分子泵-机械泵"双级泵真空系统来进行零点调整的方法。
This paper first analyzes the phenomenon of zero drift causedby the influence of structure, operation environment and artificial debugging on the absolute pressure transmitter, and then studies the linear error produced by zero adjustment of different range of absolute pressure transmitter. Finally, aiming at the linear problem caused by the lack of vacuum degree in the calibration process of small range absolute pressure transmitter, a method of zero adjustment is put forward, which uses "molecular pump-mechanical pump" two stage pump vacuum system.