灭菌参数检测在灭菌器新装、移位、大修中的应用
详细信息    查看全文 | 推荐本文 |
  • 英文篇名:Application of sterilization parameter detection in the new installation,displacement and major repair of sterilizers
  • 作者:吴佳伟 ; 杨春艳 ; 张俊明 ; 张鹏
  • 英文作者:WU Jia-wei;YANG Chun-yan;ZHANG Jun-ming;ZHANG Peng;Central Sterile Supply Department,Fifth Medical Center,Chinese PLA General Hospital;Medical Engineering Section,Fifth Medical Center,Chinese PLA General Hospital;
  • 关键词:温度压力检测仪 ; 灭菌器 ; 灭菌参数 ; 6类PCD
  • 英文关键词:temperature and pressure detector;;sterilizer;;sterilization parameter;;PCD(type 6)
  • 中文刊名:GRKZ
  • 英文刊名:Chinese Journal of Infection Control
  • 机构:解放军总医院第五医学中心消毒供应中心;解放军总医院第五医学中心医学工程科;
  • 出版日期:2019-07-28
  • 出版单位:中国感染控制杂志
  • 年:2019
  • 期:v.18
  • 语种:中文;
  • 页:GRKZ201907014
  • 页数:4
  • CN:07
  • ISSN:43-1390/R
  • 分类号:80-83
摘要
目的检测移位后脉动预真空灭菌器的运行状态是否正常。方法采用温度压力检测仪和化学、物理、生物监测方法检测灭菌器的运行状态,对检测不合格灭菌器进行维修,监测移装后灭菌器运行状态直至各项指标合格。结果移装的4台灭菌器中,2、3、4号灭菌器的运行状态良好,1号灭菌器存在问题。移装前1号灭菌器物理监测、生物监测和管腔PCD检测合格,6类PCD、B-D测试发现灭菌器故障,温度压力检测仪检测也不合格。灭菌器厂家和温度压力检测厂家工程师共同查找原因,经过3次报修后,最后通过提高灭菌器的灭菌温度和压力,灭菌器检测参数明显改善且合格,最终将1号灭菌器调整至最佳运行状态。结论除常规监测外,依据标准WS310.3—2016使用温度压力检测仪对灭菌器进行温度、压力和时间等参数进行检测是十分必要的,同时新装、移位、大修后同样应使用温度压力检测仪对灭菌器进行检测。
        Objective To check whether the status of performance operation of pre-vacuum team sterilizers is normal after displacement.Methods The temperature and pressure detector as well as chemical,physical,and biological monitoring methods were used to evaluate the operating status of sterilizers,the unqualified sterilizers were repaired and operation status of sterilizers after loading was monitored until all indicators shown as qualified.Results Among the four sterilizers that were moved,sterilizers No.2,3,and 4 were in good working condition,while sterilizer No.1 had problems.The physical,biological monitoring,and lumen process challenge device(PCD)detection in sterilizer No.1 before loading were qualified,the sterilizer fault was found by PCD(type 6)and B-D tests,temperature and pressure detector failed to pass the test.The potential causes of failures were examined by the engineers from the manufacturers of sterilizer as well as temperature and pressure detector.After three repairs,sterilization temperature and pressure of sterilizers were improved,sterilizer detection parameters were significantly improved and became qualified,the No.1 sterilizer was finally adjusted to the optimum operating state.Conclusion In addition to routine monitoring,it is necessary to use temperature and pressure detector to measure the temperature,pressure,and time of sterilizer according to the Standard WS 310.3-2016.At the same time,it is also recommended to use the temperature and pressure detector as mandatory procedure after each new installation,displacement and overhaul of the sterilizer.
引文
[1]李秋莲.新启用的脉动真空压力蒸汽灭菌器的质量监测[J].中华医院感染学杂志,2002,12(6):451,450.
    [2]张青,张金涛,高宏,等.医用脉动真空高压蒸汽灭菌设备的验证[J].中华医院感染学杂志,2008,18(4):529-531.
    [3]中华人民共和国国家卫生和计划生育委员会.医院消毒供应中心第3部分:清洗消毒及灭菌效果监测标准:WS 310.3—2016[S].北京,2016.
    [4]王旭,张剑,张青,等.应用温度压力检测仪对压力蒸汽灭菌程序进行检测分析[J].中国消毒学杂志,2018,35(8):614-615.
    [5]王旭,张青,张剑,等.两种方法对压力蒸汽灭菌器灭菌参数检测效果比较[J].中国消毒学杂志,2018,35(2):148-149.
    [6]Puttaiah R,Griggs J,D’Onofrio M.A preliminary evaluation of a reusable digital sterilization indicator prototype[J].J Contemp Dent Pract,2014,15(5):626-635.
    [7]Uetera Y,Shigematsu H,Baba Z,et al.Role of the wireless thermologger system in the management of the autoclave unequipped with sensor ports[J].PDA J Pharm Sci Technol,2003,57(6):404-409.
    [8]中华人民共和国国家质量监督检验检疫总局,中国国家标准化管理委员会.大型蒸汽灭菌器技术要求自动控制型:GB8599-2008[S].北京,2008.
    [9]林萍,马玉兰.消毒供应中心外来医疗器械灭菌湿包现象分析与改进[J].护理实践与研究,2016,13(5):131-132.
    [10]曹原,张流波.压力蒸汽灭菌过程验证装置(PCD)的进展[J].中国护理管理,2008,8(5):34-37.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700