Comparison of Atmospheric Pressure Ionization Gas Chromatography-Triple Quadrupole Mass Spectrometry to Traditional High-Resolution Mass Spectrometry for the Identification and Quantification of Halog
详细信息    查看全文
文摘
The goal of this study was to qualify gas chromatography coupled to atmospheric pressure ionization tandem mass spectrometry (APGC-MS/MS) as a reliable and valid technique for analysis of halogenated dioxins and furans that could be used in place of more traditional gas chromatography coupled to high-resolution mass spectrometry (GC-HRMS) analysis. A direct comparison of the two instrumental techniques was performed. APGC-MS/MS system sensitivity was demonstrated to be on the single femtogram level. The APGC-MS/MS analysis also demonstrated method detection limits (MDLs) in both sediment and fish that were 2鈥?8 times lower than those determined for the GC-HRMS. Inlet conditions were established to prevent issues with sample carry-over, due largely to the enhanced sensitivity of this technique. Additionally, this work utilized direct injection for sample introduction through the split/splittless inlet. Finally, quantification of both sediment and fish certified reference materials were directly compared between the APGC-MS/MS and GC-HRMS. The APGC-MS/MS performed similarly to, if not better than, the GC-HRMS instrument in the analysis of these samples. This data is intended to substantiate APGC-MS/MS as a comparable technique to GC-HRMS for the analysis of dioxins and furans.