Ex聽Vivo Characterization of Canine Liver Tissue Viscoelasticity after High-Intensity Focused Ultrasound Ablation
详细信息    查看全文
文摘
The potential of elasticity imaging to detect high-intensity focused ultrasound (HIFU) lesions on the basis of their distinct biomechanical properties is promising. However, information on the quantitative mechanical properties of the tissue and the optimal intensity at which to determine the best contrast parameters is scarce. In this study, fresh canine livers were ablated using combinations of ISPTA intensities of 5.55, 7.16 and 9.07聽kW/cm2 and durations of 10 and 30 s ex聽vivo, resulting in six groups of ablated tissues. Biopsy samples were then interrogated using dynamic shear mechanical testing within the range of 0.1-10聽Hz to characterize the tissue's post-ablation viscoelastic properties. All mechanical parameters were found to be frequency dependent. Compared with unablated cases, all six groups of ablated tissues had statistically significant higher complex shear modulus and shear viscosity. However, among the ablated groups, both complex shear modulus and shear viscosity were found to monotonically increase in groups 1-4 (5.55聽kW/cm2 for 10 s, 7.16聽kW/cm2 for 10 s, 9.07聽kW/cm2 for 10 s, and 5.55聽kW/cm2 for 30 s, respectively), but to decrease in groups 5 and 6 (7.16聽kW/cm2 for 30 s, and 9.07聽kW/cm2 for 30 s, respectively). For groups 5 and 6, the temperature was expected to exceed the boiling point, and therefore, the decreased stiffening could be due to the compromised integrity of the tissue microstructure. Future studies will entail estimation tissue mechanical properties in聽vivo and perform real-time monitoring of tissue alterations during ablation.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700